Once developed a model with the math treatments we consider adequate, and using the Calibration Samples for Instrument 1 (following with the Shootout_2002 tutorial), the idea is to check if that model performs fine with Instrument 2 for exactly the same samples. A bias is expected, because even being the same model of instrument, differences in the hardware components, optics, alignments,…, apart from some others are the cause of this bias.
Some time ago I develop a function to monitor, to plot and obtain the statistics necessaries to take some decisions as if the bias or slope should be adjusted, to check for outliers of high residuals,.....

 
It is clear that we have 5 outliers of high residuals that the literature about this set considers that must be removed, so the error RMSEP will decrease.
Due to the high RMSEP of the model used for this monitor (RMSEP=4.33 using 4 terms and with all the samples), the bias must be quite high in order to be consider that it should be adjusted. This error is more than 3 times the Lab error.

So the statistics are:


monitor14(monit.tr2[,2],monit.tr2[,1],155,4,0.95,4.33)
Where 0.95 is the confidence interval and 4.33 the CV error of the model using 4 terms


N Validation Samples  = 155 
N Calibration Samples = 155 
N Calibration Terms   = 4 
------------------------------------- 
RMSEP    : 4.942 
Bias     : -2.509 
SEP      : 4.272 
UECLs    : 4.951 
***SEP is bellow BCLs (O.K)***
Corr     : 0.9811 
RSQ      : 0.9626 
Slope    : 0.9813 
Intercept: 6.122 
RER      : 19.9   Fair 
RPD      : 5.146   Good 
BCL(+/-): 0.6778 
***Bias adjustment in not necessary***
Residual Std Dev is : 4.266 
***Slope adjustment in not necessary***

We can see how the SEP (error corrected by the bias) is similar to the error of 
the model, so a bias adjustment will help to transfer the model from instrument 1
to Instrument 2.
I will remove the 5 samples and come back with the results.