31 dic 2014

Searching / Archiving samples in ISI Scan

All the samples in ISI Scan are stored in the "data" folder in the ISI Scan directory. Keeping the Access data base, and  files included into that data folder is important to run the ISI Scan fast and secure.
So it is important to make backups regularly and clean the database at the same time. All this is configures at the System Profile window.

 
In this window, we configure the samples the samples which stay in the database and the samples which are exported or deleted definitively.
Spectra of samples with Lab Value attached are always kept in ISI Scan , so they can be exported as CAL files or used for Monitor purposes.
Read the ISI Scan Help section of System profile for a better understanding of the procedure.
After the Backup, a window appears searching for samples to archive:


Samples are archived in a new folder called Archive inside the Backup folder. Into this Archive folder there are new folders with the name of the products and inside this folders are the NIR files with the spectra, and the ANL or CSV files with the results. So you can export this files to Win ISI to work with them.

 
If the samples stay in the Data folder of ISI Scan and are not yet archived, we can search them clicking with the mouse on Products and selecting "Search". A new window appears where we have filter options, once searched the samples will go to the "Selected Samples" folder.
 
 

17 dic 2014

2 ways to draw a spectra set with plot 3D

Reading the article "Fifty ways to draw a volcano using package plot3D",
I wanted to test it with spectra (in this case treated with SNV). The result looks nice, but I have to practice more with it.
It seems that there are nice applications to use these plots in Chemometrics for tutorials, so more ideas are coming how to use them.

persp3D(z = X2.val_snv, facets = FALSE, col = "darkblue")
> persp3D(z = X2.val_snv) 

14 dic 2014

Plots to check the STD

 
Spectra plots also will help us to understand how the STD correct the spectra, looking to the spectra and study the patterns we can see for wich samples works better or worse. We expect to see as much random noise as possible.
These plots are the validation samples of the Shootout 2002, with a standardization developed with a factor matrix multiplied to the unstandardized validation spectra. As said in previous posts, 8 samples were selected, but other samples could give (or not) similar plots.
Selection of samples is an important task. What is clear is that with the STD applied we get much better statistics as we saw in the post: "Standardizing the spectra (Shootout 2002)".
 
 
 

 

13 dic 2014

Plots to prepare the STD

In the previous post 8 samples were selected for the STD.
First figure shows the selected samples for the STD scanned (raw spectra) in Instrument 1 and 2, and the differences between them.
 
We can see the same procedure, but in this case the comparative is done with math-treatments (2º Derivative + MSC). As expected much more peaks in the spectra of the difference:
 
In a recent article from Mark Westerhaus to NIR News, he advice about the importance to check and review these plots and its shape, in order to apply the best standardization (single or multiple).
Comparing with more details the spectra, there are differences in the instruments not only in the photometric scale, also in the wavelength positions and in the bandwidth, but this differences are not constant all along the wavelength axis.
Actually manufacturers are improving the instruments to match each other, specially in the wavelength axis where more complex agorithms for correction are needed.