There
are some occasions where we have to transfer equations or databases from one
instrument to other and both instruments have a different sample presentation
system, so there is no possibility to scan the same sample in the same cuvette.
In this case we have to make some repacking form one cuvette to the other at
the same time that we scan the sample on the correspondent instrument.
It
is very important that the sample must be as well homogenized as possible
between repacks. As you make more repacks there is a higher possibility that
the same sample is presented to both instruments so we can make a better standardization.
But
which sample must be chosen for the standardization and to make the repacks.
One option is to use a sample with an spectra close to the center of the spectral
population, in this case we don´t look to their value in moisture , protein and
so on. Later when we evaluate the statistics of the transfer we van see that
the standardization works fine in a certain range of the moisture range and not
so well in other ranges, so we have to consider also the values of the moisture
content of the sample we use for the standardization.
In
this case we compare the results of two instrument standardized without taking
care of the moisture content of the sample/s used for the standardization, and
we can see that in the upper range there is a difference and the monitor function
recommends a slope adjustment. We can adjust the slope after the
standardization but the idea would be to try other standardization using
samples in the moisture range and to check a better performance doing this monitor validation again.
No hay comentarios:
Publicar un comentario